15.3 Transport and storage conditions
When transporting or storing the remOVE DC Impulse, please make sure that the
following environmental requirements are met:
Environmental factors
Storage
Transportation
Temperature
0 °C to +50 °C
-20 °C to +50 °C
Relative humidity
0 % to 90 %, RH
non-condensing
0 % to 90 %, RH
non-condensing
Air pressure
500 to 1060 hPa
500 to 1060 hPa
16 Maintenance / repair
16.1 In general
After every use of the remOVE DC Impulse and its accessories, please check for
damages or defects. Pay special attention to intact isolation of all cords and ca-
bles.
Never use a damaged remOVE DC Impulse or damaged accessories. Immedi-
ately replace defective accessories.
Make sure that a safety inspection of the remOVE DC Impulse is performed an-
nually.
16.2 Safety inspection
Safety inspections have to be performed annually.
Consider that national regulations might call for more frequent safety inspections
and make sure to have the inspections performed accordingly. When performing
a safety inspection, please make sure all national requirements and rules are met.
Safety inspections of the device and its accessories may only be performed by
qualified personnel with all required knowledge and experience, who are author-
ized to perform safety inspections without supervision.
The inspector documents all testing results and measurements according to the
printable inspection record (see appendix of the instructions for use).
If considerable deviations or abnormalities are recorded, please contact the man-
ufacturer.