English; Version 08 | 2019-04-24
Multiple use of the remOVE DC Impulse
The remOVE DC Impulse is designed for multiple use. In doing so,
special cleaning measures and storage conditions are to be observed
(see chapter 14 and section 15.3).
Damage to the remOVE DC Impulse
Prior to using the remOVE DC Impulse, please check that all components are
present and that there are no defects. Incomplete or faulty components must be
replaced immediately. Faulty or missing components may result in the product
malfunctioning. For example, flawed insulation on power supply lines may lead
to electrocution of the patient and/or user.
Use of the device in the hospital environment
The characteristics of this device determined by emissions allow its use in hos-
pitals (CISPR 11, class A). When used in a home environment (for which class
B is usually required according to CISPR 11), this device may not offer sufficient
protection from radio services. If necessary, the user must take corrective
measures such as conversion or realignment of the device.
Use in the vicinity of active HF surgery devices
Do not use in the vicinity of active HF surgery devices. Active HF surgery devices
may not be used when the pulse current is in use.
Use in the vicinity of electrical devices
Use of the remOVE DC Impulse in the immediate vicinity of or
stacked on other electrical devices should be avoided as this may
lead to malfunctioning.
Use of the accessories provided
The use of components, transducers (mains adapters) and cables other than
those specified or provided by the manufacturer of this device can result in in-
creased transient electromagnetic emissions or reduced electromagnetic im-
munity of the devices and lead to malfunctioning.
Use in the vicinity of HF communication devices
Portable HF communication devices (e.g. mobile phones, devices with radio
function), including their accessories such as aerial leads and external aerials,
should be at least 30 cm away from the components indicated by the manufac-
turer and the cables of the remOVE system. Failure to comply may lead to a drop
in product performance.
MRI compatibility of the remOVE DC Impulse
The remOVE DC Impulse is not MRI compatible. It must not be used in an MRI
environment.