EasyManua.ls Logo

Ovesco remOVE DC Impulse - Page 24

Ovesco remOVE DC Impulse
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24 / 32
Instructions for use
remOVE DC Impulse 400.01
English; Version 08 | 2019-04-24
Power input
100V230 V
Power supply voltage
220V230 V
100 V120 V
Power consumption in standby mode once fully
charged
6 W / 17 VA
6 W / 10 VA
Power consumption in standby mode
60 mA
90 mA
Max. power consumption
50 W / 55 VA
50 W / 50 VA
Mains fuse
T 0.8 A H 250 V
T 0.8 A H 250 V
Line frequency
50 / 60 Hz
50 / 60 Hz
Measurements and weight
Measurements of product
340 x 340 x 110 mm
Net weight
5.0 kg
Type of packaging / measurements
Carton 400 x 400 x 300 mm
Gross weight
7.5 kg
Operating conditions
State of charge
Indicated by green flashing LEDs on status display.
Ready for use initial
Indicated by three short acoustic signals in quick succession
and continuous glowing of all LEDs on status display in
green once operating voltage is reached. Initial charge when
device energy storage is completely discharged takes ca. 10
minutes.
Ready for use signal
Indicated by three short acoustic signals in quick succession
once operating voltage is reached or 6 seconds after DC
pulse.
Ready for use visual
Indicated by all LEDs on the status display being lit in green.
On/Off signal
Yes. Green LED display above on/off switch.
Overheating protection
Yes.
Error message
Yes. All LEDs are flashing orange on status display.
Documentation
Log and record of number of uses
Yes. Readout by qualified personnel only.
Specifications
Instrument type remOVE DC Cutter
Bipolar / direct current
One-time duration of impulse during activa-
tion
65 ms
Output current remOVE DC Cutter
165 A
Output voltage remOVE DC Cutter
1.32.0 V
Interval between activations
6 s