Do Self Cal (Self-Calibration)
The self-calibration procedure can improve the accuracy of the oscilloscope under the
ambient temperature to the greatest extent. If the change of the ambient temperature is
up to or exceeds 5℃, the self-calibration procedure should be executed to obtain the
highest level of accuracy.
Before executing the self-calibration procedure, disconnect all probes or wires from the
input connector. Push the Utility button, select Function in the right menu, the function
menu will display at the left, select Adjust. If everything is ready, select Self Cal in the