EasyManua.ls Logo

Pace MBT250 - Automatic Power down; Channel Selection; Temperature Limits

Pace MBT250
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
AUTOMATIC POWER DOWN
10. The Automatic Power Down feature operates when (and only when) the Automatic Temperature
Setback feature is enabled. No additional steps are necessary. For example, power to all channels is
turned off 90 minutes after the last Active Channels Tip Temperature is set back. For additional
information on this feature, refer to the "Automatic Power Down" section of this manual.
CHANNEL SELECTION
11. The CH 1 LED is now on signifying that Channel 1 is ready for calibration. Perform steps 12 through 21
to calibrate. Change channels as directed and repeat these steps for each channel.
TEMPERATURE LIMITS
NOTE
All temperature limits are entered and stored in system memory in
degrees F.
A) LOWER TEMPERATURE LIMIT SETPOINT
12. Press and release the TIP SET Key. The Digital Readout will now display "L-X" (X = 1-9). This is the
stored value of the Lower Temperature Limit in increments of 100°F. For example, "L-5" is displayed, the
Lower Limit is 500°F.
13. Press Scroll Keys as necessary to increase (Scroll Up Key) or decrease (Scroll Down Key) the Lower
Temperature Limit value.
39

Related product manuals