Main Menu (continued)
Performing AUTO CALIBRATION
1.
Turn the rotary knob <PICTURE> in the [SYSTEM CON-
FIG] menu to select the [CALIBRATION] submenu and
press the rotary knob <PICTURE>.
2. Turn the rotary knob <PICTURE> to select the [AUTO
CALIBRATION] submenu and press the rotary knob
<PICTURE>.
MENU EXIT
RETURN
RETURNSEL. ENTER
[AUTO CALIBRATION]
YES
NO
3. Select [YES] and press the rotary knob <PICTURE>.
This starts calibration.
MENUEXIT
RETURN
RETURN
[AUTO CALIBRATION]
xxxxxxxxx
xxxxxxxxx indicates that the following message with the
meaning listed below appears.
EXECUTING: Operation in progress
COMPLETE: Operation completed
INCOMPLETE: Operation incomplete (Check the
RS-232C connection or display color
analyzer connection.)
4. Turn the power off and then on after performing auto
calibration.
RESET operation
1.
Turn the rotary knob <PICTURE> in the [SYSTEM CON-
FIG] menu to select the [CALIBRATION] submenu and
press the rotary knob <PICTURE>.
2. Turn the rotary knob <PICTURE> to select the [RESET]
submenu and press the rotary knob <PICTURE>.
MENU EXIT
RETURN
RETURNSEL. ENTER
[RESET]
YES
NO
3. Select [YES] and press the rotary knob <PICTURE>.
This starts calibration.
MENUEXIT
RETURN
RETURN
[RESET]
xxxxxxxxx
xxxxxxxxx indication.
COMPLETE: Operation completed
4. Turn the power off and then on after performing auto
calibration.