5. Insert the hose that you removed in the 2. to the T-shape
joint and the other end to the 2-row joint.
6. Push the On/off button and push the Leg massage button.
7. Check the pressures at three levels by the intensity
adjustment of the Controller. If the pressure values are in
the range of the tolerance, the Air bag and the Pump unit
are normal.
If the pressure values are low, check the pressure of the
Pump unit itself and judge the Air bag and the Pump unit.
If the Pump unit is also normal, it is assumed that there is
something wrong with the Air bag or the Electromagnetic
valves.
40
EP1260-W7