Panasonic MCO-170AICUVH Series - Parameter Setting Details

Panasonic MCO-170AICUVH Series
94 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33
3. Input each parameter. Press Apply key to save
the input value. The display returns to the Menu
screen.
Each parameter setting
Temperature: Set value of chamber temperature.
Settable range: 0.0
o
C50.0
o
C, factory setting: 37.0
o
C
CO
2
Density: Set value of chamber CO
2
density. Settable range: 0.0 %20.0 %, factory setting: 0.0 %
High Limit: The high limit temperature alarm is different from the Automatic set temperature alarm (page
37), and it is independent temperature alarm. In case the chamber temperature exceeds the temperature
of the high limit temperature alarm, this alarm is activated.
Settable range: 20.0
o
C53.0
o
C, factory setting: 53.0
o
C
Auto-lock: Refer to page 6769.
Refer to page 7880 for detail of each alarm.
Note:
When operating the incubator for the first time or after not using it for an extended period of time, operate
it for at least about 4 hours until the chamber temperature and the CO
2
sensor are stable after setting the
chamber temperature to the desired temperature and setting the CO
2
density to 0 %. Then change the
setting to the desired CO
2
density.
Set the temperature of the high limit temperature alarm after the chamber temperature is stable at the set
value.
Set the high limit temperature alarm to at least 5
o
C higher than the chamber set temperature.
4. On the Menu screen, press Back key to return to the Top screen.