ALARM PARAMETERS
1. Press Menu key to lead the Menu screen.
2. Press Tools #1 key to lead the Tools #1 screen.
3. On the Tools #1 screen,
Press Alarm Setting #1 key to lead the Alarm
Setting #1 screen, it is possible to set automatic set
temperature alarm and automatic set CO
2
density
alarm (go to procedure 4).
Press Alarm Setting #2 key to lead the Alarm
Setting #2 screen, it is possible to set alarm delay,
door alarm delay, ring back and remote alarm (go
to procedure 5).
4. On the Alarm Setting #1 screen, input each
parameter. Press Apply key to save the input value.
The display returns to the Tools #1 screen.
Each parameter setting
Tem p . A l a r m :
When the chamber temperature exceeds the scope,
the set temperature ± the set value of Automatic set
temperature alarm, the alarm is activated. Settable
range: ±1.0
o
C±5.0
o
C, factory setting: ±1.0
o
C
CO
2
Alarm:
When the chamber CO
2
density exceeds the scope, the set CO
2
density ± the set value of Automatic set
CO
2
density alarm, the alarm is activated. Settable range: ±0.5 %±5.0 %, factory setting: ±1.0 %