9. Set the humidifying pan cover, the humidifying pan removed in the procedure 1 (Fig.3)
Note: H
2
O
2
decontamination can be performed only for the chamber and inner attachments with standard
specifications, not for any other objects.
10. Make sure that the duct, the fan and the H
2
O
2
generator are securely attached (refer to Fig. 1 to 3 on
page 83). After that, close the inner door and the outer door.
11. Press H
2
O
2
key for 3 seconds to lead the H
2
O
2
Decontamination screen.
Note:
When the model does not have H
2
O
2
decontamination function, H
2
O
2
key is not
displayed on the Top screen.
When key lock is ON, Password input window is
opened and input of the password is required.
Refer to page 35.
12. On the H
2
O
2
Decontamination Step1 screen,
the system check starts automatically. If the system
is normal, the display leads the H
2
O
2
Decontamination Step2 screen. When the system
is abnormal, refer to Table 8 on page 80.
Fig. 3
Front
Humidifying
pan cover
Humidifying pan
H
2
O
2
generator