65
Step7 is the process to resolve H
2
O
2
vapor by UV
lamp.
Note: Step5 and Step6 do not exist.
14. When the H
2
O
2
decontamination is completed,
the H
2
O
2
Decontamination Step8 screen is
displayed.
Open the outer and inner doors. Disconnect
connector of the chamber, cable and H
2
O
2
generator, and remove H
2
O
2
generator and cable
from the chamber.
Note: When doing the above work, put on
protection glasses and rubber gloves.
15. On the H
2
O
2
Decontamination Step8 screen,
press OK key to return to the Top screen.
Note: When H
2
O
2
generator is connected with
connector of the chamber by cable, OK key is not
workable.
16. Dilute the remaining H
2
O
2
reagent in the H
2
O
2
generator with a large volume of water and dispose of it.
Rinse and wash the H
2
O
2
generator with distilled water. Then keep the H
2
O
2
generator in a clean
environment outside of the chamber.
Note: Do not wash either the inside or outside of the H
2
O
2
generator with alcohol.
17. After H
2
O
2
decontamination, surplus H
2
O
2
liquid will remain at the
bottom of the chamber and in the bottom part of the H
2
O
2
generator
duct. This solution contains H
2
O
2
at a low density, so put on protective
glasses and rubber gloves and wipe it up with a non-woven cloth.
18. Ventilate the chamber sufficiently and place all the attachments
back into the chamber.
Note: After H
2
O
2
decontamination, cover the connector on the
chamber side with the connector cap deeply. (Fig. 4)
Fig. 4
Connector cap