STD GAS AUTO CALIBRATION KIT (OPTION)
9. On the STD Gas Calibration Step3 screen, CO
2
density calibration starts. Calibration go to Step5
(Procedure 10) automatically.
10. After completion of CO
2
density calibration,
display leads the STD Gas Calibration Step5
screen. CO
2
incubator returns to the normal
operation.
11. On the STD Gas Calibration Step5 screen,
press OK key to return to the Tools #1 screen. On
the Tools #1 screen, press Top key to return to the
Top screen.