Panasonic MCO-170AICUVH Series - Alarms for Specific Functions

Panasonic MCO-170AICUVH Series
94 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
79
Alarm o
r
safety
function
Conditions Display Buzzer
Remote
alarm
Safety
operation
UV lamp failure
(MCO-170AICUVH/MCO-170AICUV,
or when an optional MCO-170UVS is
installed)
The UV lamp burns out
“Err18: UV Lamp Abnormal” is displayed
in the message display field.
Intermittent
tone
ON
-----
New UV lamp
replacement
(MCO-170AICUVH/MCO-170AICUV,
or when an optional MCO-170UVS is
installed)
The accumulated ON time reaches
approx. 5,000 h
“Warning: UV Bulb Life” is displayed in
the message display field.
----- ----- -----
Communication
error
When communication between LCD
touch panel and control substrate is
died out or unstable.
“Err56: Communication Failure” is
displayed in the message display field.
----- ----- -----
Table 57 show the behavior of the alarm (buzzer) and Ring Back function when pressing Buzzer key.
Table 5 In the cases of other than table 6 or table 7.
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator Remote Alarm
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
OFF
(Alarm is not
canceled)
ON
ON
ON
(Under continuation)
OFF OFF
OFF: Interlock
with Buzzer key
ON ON
OFF (Alarm is
not canceled)
ON
OFF OFF OFF
Note: Resolve the cause of the alarm in reference to page 78-80 because the alarm itself is not
deactivated by pressing Buzzer key
Table 6 In the cases of high limit temperature alarm or Err38 (The outer door opens during H
2
O
2
decontamination; refer to next page)
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator Remote Alarm
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
ON
ON
(Under continuation)
ON
(Continue)
ON
(Under continuation)
OFF
OFF: Interlock
with Buzzer key
ON
OFF
Note: Close the outer door when Err38 is activated.
Table 7 In the cases of Err01 (CO
2
gas cylinder empty), Err11, 12 (CO
2
sensor error), Err18 (UV
lamp failure) or door alarm*.
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator Remote Alarm setting
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
OFF
(Alarm is
canceled)
OFF
(Alarm is already
canceled)
OFF
(Alarm is
canceled*
2
)
OFF
(Alarm is already
canceled*
2
)
OFF
OFF: Interlock
with Buzzer key
ON
OFF
*2: In the door alarm, the remote alarm does not work.
Note: Err01 recovery is achieved automatically by correctly connecting the new CO
2
gas cylinder and
pressing the Buzzer key to stop the buzzer. In addition, when the optional MCO-21GC is installed and the
gas supply is switched to the reserve gas cylinder, press the Buzzer key and replace the gas cylinder.