MECHANICAL INSTALLATION
A. Fit one of the ½ ˝ball valves supplied to the compressed air inlet port on the generator and attach the compressed air supply to this ball
valve. Ensure that the valve is in the closed position.
B. Fit another of the ½ ˝ball valves supplied to the port marked “To Bu er vessel”. Install ½ ” NB / 16mm ID piping between the ball valve and
the bu er vessel inlet port. It is recommended that a ball valve (not supplied) be installed at the inlet to the bu er vessel to allow it to be
isolated during maintenance.
C. Fit the remaining ½ ˝ball valve to the port marked “From Bu er vessel”. Install ½ ” NB / 16mm ID piping between the ball valve and the
outlet port of the bu er vessel.  e AOP010  lter provided should be installed in this line. Follow the installation instructions provided
with the  lter taking note of the direction of  ow. It is recommended that a ball valve (not supplied) be installed at the outlet of the bu er
vessel to allow it to be isolated during maintenance.
D. Fit the 3-way ball valve supplied to the port marked “Nitrogen Outlet”. Connect this ball valve to the application using ½ ” NB / 16mm ID
piping. is piping must be solid and non-porous to minimise the ingress of oxygen.
Note.  e nitrogen bu er vessel must be rated to at least the maximum operating pressure of the generator and must be  tted with a
suitable pressure gauge and pressure relief valve.
A
C
B
D
16
N Compact Nitrogen Gas Generator - User Guide. ©2020.