SciCon Disposable Conductivity / Temperature Sensors:
CONNECTOR TYPE Max FLOW RATE Max PRESSURE NOMINAL CELL VOLUME
Luer
1 liter/min 60 psi 80 µl
3/8" Barb
8 liters/min 60 psi 0.5 ml
1/2" Barb
17 liters/min 60 psi 0.8 ml
3/4" TC
31 liters/min 60 psi 1.2 ml
1” Ladish TC
60 liters/min 60 psi 2.3 ml
Specifications for all SciCon™ Conductivity Flow Thru Sensors
Material, Fluid Contact:
Polysulfone (PS) w/Gold Electrodes; Medical Grade meets USP Class VI
and FDA 21 CFR 177.1520. All Wetted Materials are made of Animal-Free
Compounds. Most Sanitizing Agents such as NaOH, Hypochlorite. Sensors
can be autoclaved..
Sensor Type:
4-electrode conductivity cell
Factory calibrated, Ready to use
Conductivity Range:
1 µS/cm to 200 mS/cm
Accuracy:
High Range:
± 0.25 mS in the 0.1 - 200 mS range;
Low Range: ± 3 μS in the 0 - 100 μS range.
Temperature Range:
0 - 50 Celsius
Temperature Probe:
Thermistor, Factory Calibrated
Temp. Accuracy:
+/- 0.5 Celsius
Sensor Microchip:
EPROM, stores Device ID, Cell Constant, Temp Offset & Factory Cal Data