SciPres Disposable Pressure Sensors:
CONNECTOR TYPE Max FLOW RATE * Max PRESSURE NOMINAL CELL VOLUM
E
Luer
1 liter/min 60 psi 80 µl
3/8" Barb
8 liters/min 60 psi 0.5 ml
1/2" Barb
17 liters/min 60 psi 0.8 ml
3/4" TC
31 liters/min 60 psi 1.2 ml
1” Ladish TC
60 liters/min 60 psi 2.3 ml
*At a 1.0 psi drop across the sensor
Specifications for all SciPres Flow Thru Pressure Sensors
Material, Fluid Contact:
Medical Grade Polysulfone meets USP Class VI and FDA 21CFR177.1520.
All Wetted Materials are made of Animal-Free Compounds. Compatible with
Most Sanitizing Agents such as NaOH, Hypochlorite. Sensors can be
autoclaved.
Sensor Type:
Medical grade silicone piezoresistive sensing element with on-chip
temperature compensation
Sensor Isolation:
Insoluble silicone dielectric gel isolates sensing element from process
solution.
Pressure Sensor Range:
0-60 psi for all sizes.
Pressure Sensor Accuracy:
+/- 0.30 psi 0-60 psi.
Pressure Sensor Resolution:
0.01 psi.
Temperature Range:
0 - 60 Celsius
Pressure Sensor Microchip:
EPROM, stores Device ID, Cell Constant, Temp Offset & Factory Cal Data
Pressure Sensor Cleaning:
Sensors may be flushed with Alcohol, 0.1M or 0.5M NaOH, dilute Bleach, or
Formaldehyde for inline sanitization several times. The Sensors may also
be autoclaved.
Sensor Connector and Cables: Lockable & Waterproof
The maximum recommended pressure for the sensors is 60 psi. If this is
exceeded, problems with leakage and functionality can occur.