31
SciTemp Disposable Temperature Sensors:
CONNECTOR TYPE Max FLOW RATE * Max Pressure
Luer
1 liter/min 60 psi
3/8" Barb
8 liters/min 60 psi
1/2" Barb
17 liters/min 60 psi
3/4" TC
31 liters/min 60 psi
1” Ladish TC
60 liters/min 60 psi
*At a 1.0 psi drop across the sensor
Specifications for all SciTemp™ Temperature Sensor Flow Thru Temperature Sensors
Material, Fluid Contact:
Natural PP, meets USP Class VI and FDA 21 CFR 177.1520
Medical Grade Polysulfone meets USP Class VI and FDA 21CFR177.1520.
All Wetted Materials are made of Animal-Free Compounds. Compatible with
Most Sanitizing Agents such as NaOH, Hypochlorite. Sensors can be
autoclaved.
Sensor Type:
Thermistor, Epoxy Coated, 2252 ohm.
Temperature Sensor Range:
-40 to +150 ° C for all sizes.
Temperature Sensor
Accuracy:
0.01° C In the 4.0 to 70.0° C Range.
Temperature Sensor
Resolution:
0.01° C.
Temperature Sensor
Microchip:
EPROM, stores Device ID & Factory Cal Data
Temperature Sensor
Cleaning:
Sensors may be flushed with Alcohol, 0.1M or 0.5M NaOH, dilute Bleach, or
Formaldehyde for inline sanitization several times. The Sensors may also
be autoclaved.
Sensor Connector and Cables: Lockable & Waterproof