SciCon Sensor Sanitization Test Protocol
Parker/Scilog DN 3023: SciCon Sanitization, 1.0 Molar NaOH @ 22 ºC and 50 ºC, 2 Hours
NOTE: Parker sensors are designed for disposable, single-use applications. However, with proper care, the
Parker sensors can be re-used repeatedly while maintaining good accurate and precision. If required, SciCon
sensors can be re-calibrated.
CAUTION: Do Not Exceed Maximum Pressure of 60 psi
Test Protocol: Six pre-calibrated (13.04 mS) SciCon Luer conductivity sensors were exposed 3 times to 1.0
molar NaOH for two hours each at 22 ºC (Group1) and at 50 ºC (Group 2). After each 2-hour exposure, the
sensors were flushed 3x with distilled water and inserted into a temperature controlled glove box. A solution was
then circulated through the sensors utilizing a peristaltic recirculation pump. The sensors were conductivity-tested
and temperature-tested at the indicated standard values. The conductivity standard of the solution was tested
using a NIST-traceable (YSI Model 30) conductivity meter. The temperature standard was measured using a
NIST traceable temperature thermometer. The sensor responses were tabulated. The original factory calibration
was maintained during the trials; no additional sensor calibrations were carried out before or during the trials. For
this test, SciCon sensors were randomly selected from Parker inventory.
Group 1: SciCon Sensor Conductivity Response after Exposure to 1.0 N NaOH Solution, Temperature
(1.0 N NaOH) at 22 º C, 2 Hours
Group 1 TRIAL 1 TRIAL 2 TRIAL 3
Temperature 22 ºC NIST SciCon NIST SciCon NIST SciCon NIST SciCon NIST SciCon NIST SciCon
Sensor ID Cond. Cond.
Temp Temp Cond. Cond. Temp Temp. Cond. Cond. Temp. Temp.
C1-240173-0308 13.04 13.10 25.64 25.4 13.04 13.14 24.70 24.3 13.04 13.10 24.99 24.8
C1-240251-0308 13.04 13.06 25.64 25.5 13.04 13.11 24.70 24.4 13.04 13.08 24.99 24.8
C1-240302-0308 13.04 13.00 25.64 25.7 13.04 13.05 24.70 24.6 13.04 13.04 24.99 25.0
Group Average. 13.04 13.05 25.64 25.5 13.04 13.10 24.7 24.4 13.04 13.07 24.99 24.9
Group SD* 0.05 0.2 0.05 0.2 0.03 0.1
SD* = Standard Deviation
Group 2: SciCon Sensor Conductivity Response after Exposure to 1.0 N NaOH Solution, Temperature
(1.0 N NaOH) at 50º C, 2 Hours.
Group 2 TRIAL 1 TRIAL 2 TRIAL 3
Temperature 50 ºC NIST SciCon NIST SciCon NIST SciCon NIST SciCon NIST SciCon NIST SciCon
Sensor ID Cond. Cond.
Temp Temp Cond. Cond. Temp Temp. Cond. Cond. Temp. Temp.
C1-240010-0208 13.04 13.02 25.64 25.7 13.04 13.12 24.70 24.6 13.04 13.10 24.99 25.0
C1-240060-0208 13.04 13.20 25.64 25.5 13.04 13.26 24.70 24.4 13.04 13.21 24.99 24.9
C1-240100-0208 13.04 12.91 25.64 25.5 13.04 12.93 24.70 24.7 13.04 12.84 24.99 25.2
Group Average: 13.04 13.04 25.64 25.6 13.04 13.10 24.7 24.6 13.04 13.05 24.99 25.0
Group SD 0.15 0.1 0.17 0.2 0.19 0.2