EasyManua.ls Logo

Partec CyFlow space - Threshold: Lower Level (L-L)

Default Icon
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Basics
Rev. 007 Date 2010-01-13 CyFlow
®
space Instrument Operating Manual 17/26
Threshold: Lower Level (L-L)
Lower Level (L-L)
Note
L-L and U-L cannot be
adjusted after the
acquisition.
After acquisition, you
may use 1P ranges or
2P polygons to gate out
undesired signals
instead.
Note
L-L values do not
directly correlate with
axis values.
Typical effect of
change
L-L = 10
L-L = 40
Adjustment
Lower Level L-L
Setting a proper lower level (L-L) avoids the
acquisition of small and unwanted background or
"noise" signals below a threshold.
The L-L value of a parameter can be increased
(decreased) by clicking into the corresponding L-L
value field in the instrument settings box and using
the Right (Left) buttons (see Fig. 4).
Start with a low L-L (e.g. 10). Increase the L-L until
no more of the small "noise" signals appear in the
histograms. Make sure not to remove signals from
particles of interest by a L-L being too high.
A L-L range of 0…999.9 corresponds to the full
histogram scale, independent of the channel
resolution actually selected.
If a leading trigger is selected, then the L-L can only
be effectively set on this trigger parameter.