Rev. 007 Date 2010-01-13 CyFlow
®
space Instrument Operating Manual 18/26
Appendix
Installation Requirements
Line Power
CyFlow
®
space Basic Unit AC 100-240 V 50-60 Hz, 250 VA max
Instrument Size
CyFlow
®
space Basic Unit 560 mm x 650 mm x 300 mm (W x D x H)
Operating Environment
The CyFlow
®
space should be placed on a solid base, e.g. a laboratory table. It must be placed horizontally.
Reduce smoke, dust, vibrations, direct sunlight and direct neighbourhood of heatings as possible. The
installation room must be well ventilated and dry.
Temperature 15-30 C
Humidity 20-85% relative (non-condensing)
Room Clean environment. Direct sun light should be avoided.
Minimum Recommended Work Space
Table dimensions and space on desktop 120 cm x 90 cm x 80 cm (W x D x H)
Fig. 7: Recommended placement of CyFlow
®
space components
CyFlow
flow cytometer
Computer
Keyboard
Mouse
Screen
Fluid
containers
Printer
CyFlow
flow cytometer
Computer
Keyboard
Mouse
Screen
Fluid
containers
Printer