Appendix
Rev. 007 Date 2010-01-13 CyFlow
®
space Instrument Operating Manual 24/26
CyFlow
®
space - Technical Specifications
Note Due to fast technological improvements, specifications herein are subject to change. For details,
please inquire information from your supplier.
1. CyFlow
®
space System
Size Flow cytometer: 560 mm x 650 mm x 300 mm (W x D x H)
Weight ca. 37 kg
Maximum sound < 70 dBA
power level
Installation/overvoltage 2/II
category
Degree of protection IP 20
Operating Environment Temperature 15-30 C
Humidity 20-85% relative (non-condensing)
Room Clean environment. Direct sun light should be avoided.
Applications Research applications, Routine and Research Immunophenotyping, DNA
Analysis, Apoptosis, Blood Cell Analysis, HIV monitoring, Leukocyte
Counting / Rare Event Analysis, Microorganism Analysis (live / dead,
others), Fermentation control, Particle Concentration Analysis, True
Volumetric Absolute Counting, Particle Size and Fluorescence Distribution
Analysis
True Volumetric Based on precise counting and mechanical fluid volume measurement. No
Absolute Counting need for reference sample or beads
Instrument Check Blue laser: Product no. 05-4018 Calibration Beads 3 µm
Product no. 05-4010 CountCheck Beads
Red laser: Product no. 05-4012 Red Flourescent Particles
UV laser: Product no. 05-4020 Calibration Beads UV
Violet laser: Product no. 05-4018 Calibration Beads 3 µm
Set-up Time 10 minutes
Parameters Up to 9 optical parameters: FSC, SSC, FL1, FL2, FL3, FL4, FL5, FL6, FL7
Time parameter
Pulse height, pulse area, pulse width for each optical parameter
Particle Size Range 0.1 µm - 50 µm (standard cuvette)
Maximum Acquisition Speed 25,000 events/sec
Acquisition Stop Time Time-, event- or volume-based
Trigger On all parameters or on specific trigger parameter, selectable by software
Data Resolution 65,536 channels (16 bit)
Service 1-3 years service contracts