Page 41
COM 2 PARAMETERS WHEN ETHERNET IS PRESENT
1P
IP ADDRESS
ETHERNET protocol IP address
Values from 0.0.0.0 to 255.255.255.255
Default: 192.168.0.201
SubnEt
SUBNET MASK
ETHERNET protocol Subnet Mask.
Values from 0.0.0.0 to 255.255.255.255
Default: 255.255.255.0
GatE
GATEWAY
ETHERNET protocol gateway.
Values from 0.0.0.0 to 255.255.255.255
Default: 192.168.0.1
Port
PORT
Communication port for ETHERNET protocol.
Values from 1 to 65535
Default: 1800
Eth.Pro..
ETHERNET COMMUNICATION PROTOCOL
Selecting communication type for Ethernet protocol.
Values that can be selected:
None: Serial communication OFF
Contin: Continuous transmission of the weight string. It can be used, for example, to drive a weight
repeater. See details in the relevant section.
on deM: When the Operator presses the relative button on the front or uses Input 2, a weight string
is sent. The command is acceptedif the weight is stable. Between two consecutive transmissions,
variation of the weight must be at least equal to 20 divisions.
Autom: A weight string is sent automatically when theweight stabilises at a value above the minimum
weight (20 divisions).
Slave: ASCII protocol. See details in the relevant section.
Modbus: Modbus TCP protocol.
Default: Slave