Page 18
INFO DISPLAY
PdAt01 When the instrument is switched ON, a display test is run followed by the identifica-
tion code and then the version of the software, in that order. These codes are to be
cited when requesting assistance
When is not in progress a programming procedure, the display shows the weight measured in kilograms.
Under certain conditions, the following messages are reported:
NOTIFICATION OF ERRORS
In operating mode, the following error codes may appear on the display.
Fixed message
P <picco> Display of peak value.
rENotE Communication with “Optimation” PC utility software.
ϩϩϩϩϩϩ Overload. The load applied to the load cells exceeds by over 9 divisions the maximum
capacity of the weighing system.
O-L Signal that the load cells are absent or outside of the measuring range mV/V.
Flashing message, alternating with the weight measured.
No CAL Calibration weight not performed)
No CoN Fieldbus network disconnected
E-F.buS Fieldbus interface absent or not working.