9. Technical specifications
9.1. Sensor
Solid-state sensor with electrolyte made of stabilized zirconium oxide, which is pre-heated
to a constant temperature of 1000K (727
o
C). The warm-up time is approx. 20 minutes,
and an LED on the front plate indicates that the sensor is ready to measure, while a +15
VDC output signal is transmitted through the 8-pole DIN connector.
9.2. Measuring range
The device can be used to measure oxygen in inert gases, argon, nitrogen and formier gas.
Front control switch and/or electrically controlled switch for measuring gas range
selection ("RANGE"):
Green
Formier 2%
8.6 - 14 mA /
2.15 - 3.5 V
500 - 10 ppmFormier gas 2%
Green
Formier 5%
8.6 - 14 mA /
2.15 - 3.5 V
500 - 10 ppmFormier gas 5%
Green
Formier 10%
8.6 - 14 mA /
2.15 - 3.5 V
500 - 10 ppmFormier gas
10%
Green
Inert
4 - 20 mA /
1 - 5 V
5000 - 1 ppmInert
Indication LEDOutput signal
I out / V out
Measuring range
SGI-3
Green
ppm
4 - 20 mA /
1 - 5 V
1000 - 1 ppmppm
Green
%
4 - 20 mA /
1 - 5 V
20,9 - 0,1 % %
Indication LEDOutput signal
I out / V out
Measuring range
OXI-3
+/- 1%0 - 20 mA /
0 - 5 V
0 - 1000 ppmppm
+/- 1%0 - 20 mA /
0 - 5 V
0 - 100 %%
AccuracyOutput signal
I out / V out
Measuring range
OXI-3.1
SGI-3/OXI-3/OXI-3.1 User manual PBI-Dansensor A/S
PBI-210247-H UK 07/11/2005 19