EasyManua.ls Logo

PBI Dansensor SGI-3 - Flow System (SGI-3 & OXI-3); Flow System (OXI-3.1); Response Time; Measuring Accuracy

Default Icon
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
+/- 1%0 - 20 mA /
0 - 5 V
0 - 1000 ppmppm
+/- 1%0 - 20 mA /
0 - 5 V
0 - 2 %%
AccuracyOutput signal
I out / V out
Measuring range
OXI-3.1, 0-2% O
2
+/- 1%0 - 20 mA /
0 - 5 V
0 - 1000 ppmppm
+/- 1%0 - 20 mA /
0 - 5 V
0 - 8 %%
AccuracyOutput signal
I out / V out
Measuring range
OXI-3.1, 0-8% O
2
9.3. Flow system, SGI-3 & OXI-3
The measuring gas is supplied continuously by a diaphragm pump with a manual on/off
switch on the front. LED indication of pump operation.
Use the smallest possible hose dimensions, ø 5/3mm.
Maximum suction length: 15 m at ø 5/3 hose dimensions
9.4. Flow system, OXI-3.1
A 5 mbar pressure difference between gas inlet and gas outlet will ensure the correct flow
rate of approx. 200 ml/min. (Min. 3 mbar (140 ml flow rate) / Max. 7 mbar (300 ml flow
rate)).
When testing the indicator, supply a gas flow of 200 ml/min at the gas inlet. Do not
connect the gas outlet when testing.
9.5. Response time
Sensor: Less than 50 milliseconds
System: Depends on the suction length
9.6. Measuring accuracy
Better than ± 1 LED.
Note: The OXI-3.1 was designed to measure oxygen in inert gases online in situations
where visual display is not required and where the output signal is linear.
9.7. Indication, SGI-3 & OXI-3
LED row consisting of 48 red LEDs showing the current oxygen content. The number of
LEDs lit increases as the oxygen content in the measuring gas drops. LED indication of:
mains voltage, pump operation, sensor function and measuring gas selected.
The SGI-3 was designed to measure oxygen in inert gases (inert - argon/nitrogen) or
formier gases. Before leaving the factory, the device was calibrated for measuring with
10%, 5% and 2% H
2
, respectively, in formier gas and N
2
/O
2
mixtures. If requsted, the
SGI-3 can be calibrated with other H
2
(hydrogen) ratios in formier gases.
PBI-Dansensor A/S SGI-3/OXI-3/OXI-3.1 User manual
20 PBI-210247-H UK 07/11/2005