EasyManua.ls Logo

PEAK INSTRUMENTS P-510 - Parameter Setup; Table of Ph Meter Parameter Setup; Manual Temperature Compensation Setting(P1); Resolution Setting(P2)

PEAK INSTRUMENTS P-510
14 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Manual PEAK INSTRUMENTS INC
Address: 16223 Park Row, Houston, TX-77084, USA. Website: www.peakii.com. Tel: +1 2819353455
7
4. Parameter setup
4.1. Table of pH meter parameter setup
4.2. Manual temperature compensation setting(P1)
4.2.1. Long press to enter set mode
Then press to enter P1.
4.2.2. Press or to change temperature. Long press these keys to
change the figures continuously. Press to confirm the change and back.
4.2.3. Press to enter next parameter setting or press to back to
measuring mode.
4.3. Resolution setting(P2).
4.3.1. Press to choose P2 in mode P1,
Then press to enter P2.
4.3.2. Press or to select resolution between 0.1 and 0.01. Press
to confirm the change and back.
Indicator
Description
Remarks
P1
Manual temperature compensation
setup
(0-99.9)
P2
pH resolution setup
0.1pH/0.01pH
P3
pH buffer setup
CH, USA, NIST
P4
Temperature unit setup
P5
Ammonia pure water compensation
setup
OFF, ON
P6
Backlight time closing setting
0-20min or 0 stands for this
function is not working
P7
Automatic shutdown time setting
0-20min or 0 stands for this
function is not working
P8
System restore setup
OFF or ON

Related product manuals