8. Plug in the ribbon cable to the transmitter board and secure the board with
thumb screws. Do not over tight the screws.
9. Power up the sensor, if disconnected, and let the sensor warm-up as indicated
in the “Start up” section in this manual.
10. Calibrate the sensor. Follow the instructions in calibration procedure in this
manual.
** Note. Detector transmitter must be Calibrated any time a new
sensor head assembly is installed or replaced.