Type
Range
(user
Selectable)
Calib Gas
Used
time (T90)
seconds
Time (T10)
seconds
Default alarm Levels /
Set Points
Electrochemical
Sensors
-4°C / 24°F 50°C / 122°F 5 ppm 10 ppm
(CO)
-20°C / -4°F 55°C / 131°F 20 ppm 50 ppm
(CO)
-20°C / -4°F 55°C / 131°F 100 ppm 150 ppm
-20°C / -4°F 55°C / 131°F 2 ppm 5 ppm
-20°C / -4°F 55°C / 131°F 200 ppm 500 ppm
(H2S)
-40°C / -40°F
65°C / 149°F
(intermittent)
10 ppm 15 ppm
(H2S)
-40°C / -40°F
65°C / 149°F
(intermittent)
50 ppm 100 ppm
-20°C / -4°F 55°C / 131°F 10 ppm 15 ppm
-20°C / -4°F 55°C / 131°F 10 ppm 15 ppm
-20°C / -4°F 55°C / 131°F 20 ppm 50 ppm
Calibrated in fresh air.
Purged with Pure Nitrogen
from 20.9% to 0
-20°C / -4°F 55°C / 131°F
(falling)
-20°C / -4°F 55°C / 131°F 5 ppm 10 ppm
-40°C / -40°F 75°C / 167°F 2 % 3.5 %
(CO2)
-40°C / -40°F 75°C / 167°F 20 % 50 %
Used
(T90) seconds
(T10) seconds
%
Default alarm Levels /
Set Points
Flammable
(Hydrocarbons)
100 % LEL
Air)
Catalytic
Bead
1 % 50 % LEL
< 25
~ 40
± 2
-40°C / -40°F 75°C / 167°F 20 % 40 %
Hydrocarbons
(Methane /
100 % LEL Infrared 1 %
50 % LEL
(methane or
propane)
< 25
~ 40
± 2
-40°C / -40°F 75°C / 167°F 20 % 40 %
Table 1. Reference Table for Target Gas Range and default parameters