1020 Mono Filler
18
4.2 Recipe parameters
These parameters correspond with the parameters in the tree structure of the device Recipe.
When using the industrial protocol register functions, each parameter can be reached using its
number.
Some parameters can be reached directly using ASCII, TP protocol, Modbus RTU, Modbus TCP,
Fins, Profibus or EtherNet/IP. The parameters for Profibus or EtherNet/IP are prefixed with PB
or EIP.
Note: when the device is rebooted or the recipe is manually changed, all recipe parameters are
changed back to the value that were last set manually in the recipe.
The amount of product that is wanted on/into or out of the
weigher.
The selection net or gross and in or out is made in the
EIP-R11
Coarse dosing stops when the setpoint minus the turnover
is reached. The dosing continues in fine mode. The
correction strength is set in the configuration menu.
Δ The remaining fine time must be more than the K.E.B.
EIP-R12
The amount of product that falls on/into the weigher after
the fine output is switched off. This value can be
automatically corrected using the inflight correction in the
EIP-R13
tolerance
The allowed tolerance of the end value under the setpoint.
The fine output will stay on until the weight is within this
range.
Δ Not used if no tolerance is selected
tolerance
The allowed tolerance of the end value above the setpoint.
An alarm will be generated until the weight is accepted by
input 2.
Δ Not used if no tolerance is selected
During coarse dosing mode this value is used for the analog
output.
During fine dosing mode this value is used for the analog
output.