24
WARNING
Failure to perform the daily safety checks and take
corrective action, when required, prior to the operation of
the equipment each day could result in injury to the resident or
operator.
WARNING
If any part of the system is not functioning properly, cease all
bathing activities until the problem is corrected by maintenance.
The system must be maintained on a scheduled basis to ensure
it is functioning properly. Failure to heed these precautions
could result in injury to the operator or resident.
CHECK THE FOLLOWING ITEMS EACH DAY BEFORE USING
YOUR CASCADE™ SIT-BATH SYSTEM 6900.
Perform the following safety checks daily for the
Cascade™ Tub:
1. High temperature limit – Press the Fill Button and turn the
Temperature Control Knob all the way to the left to the highest
setting. Test the fill and shower water temperature to ensure
that it reaches no more than 115 degrees Fahrenheit (lower if
required by local regulations).
2. Seal Wear - Open the door and check to see that the seal is not
cracked or peeling away from the door.
3. Self-Draining Trap - If water is draining regularly from the self
draining trap, the 1-1/2 check valve may need to be cleaned or
replaced.
3. Swivel Lift and Seat Belt – Check the condition of the lift
and belt(s) for signs of excessive wear.
Daily Safety Checklist for Cascade™ Tub and
Reservoir