10
2-2 WATER BOTTLE ASSEMBLY, MODEL OS-H4
No. NAME FUNCTION
1 Bottle holds sterile water for procedure. (use up to 2/3’s full)
2 Water Bottle Cap Assembly
must be firmly secured to bottle to prevent air leakage. Do NOT overtighten the bottle cap.
3 Air Pipe Stem inserts into video processor water bottle receptacle.
4 Air/Water Hose contains (2) independent tubes - 1 for air, 1 for water.
5 Air/Water Connector inserts into Air/Water socket of endoscope umbilical connector.
6 Air/Water- Drain Lever must be set to upright (A/W) position for delivery of air and water.
7 Water Feeding Stem channel for water to be displaced from bottle and into scope.
- If the water bottle assembly has been handled roughly, the water feeding tube inside the Air/Water hose may be
disconnected at the A/W connector for the endoscope. To test, remove the cap assembly and using a syringe, inject water into
the water feeding stem. If the water comes out of both the center hole of the A/W connector and the series of holes around the
center hole, the water feeding tube is disconnected. Use another water bottle assembly.