23
Competency
Competency Test Key
1. Identify the following controls and their function.
HF-Control rate of breaths per minute
Amplitude-Controls pressure change of breath
Mean Airway Pressure-Allows user to switch between P
aw
and
Amplitude reading on manometer.
2. Explain in your own words how the Phasitron® functions.
See manual
3. If transporting a patient in HF mode name two (2) vent changes that would
decrease CO
2.
Decrease HF Rate
Increase Amplitude
4. How is mean airway pressure controlled during HF ventilation?
PEEP Valve
Amplitude
HF Rate
5. What is the ventilator classification of the TXP®-2D?
Pressure limited, Time cycled, HF, Flow interrupter
6. What NO system is recommended for the TXP®-2D?
Aeronox®
7. If the patient breathes spontaneously during NO delivery will concentrations be
maintained?
No, Gas will be entrained through the entrainment port diluting the
NO concentration.