PERET GmbH/S.r.L www.peret.it info@peret.it
In Color mode you can check the relationship between fluorescent
responses in red, green and blue by clicking through the colors pressing the
‘next’ button. In Intensity mode you can toggle between near UV and far UV
Intensity by pressing the right button.
Measure an average solid reference [C 2]..[C 10]
If the reference measurement counter is greater than 1, there have to be
taken as many solid reference readings as specified in the settings.
Position the FLUO with the aperture onto the fluorescent reference
surface, for example onto the substrate with fluorescent ink. Press down the
‘up’ key and push the device to the front until [C 1] appears in the LCD.
Now measure the next reference again with the middle key depressed.
Continue to measure references until the number of references specified in
the settings is reached. Once all reference measurements are taken, the
device will calculate the average and set this average to 100%
Measure samples in color mode [CoL]
Position the FLUO with the aperture onto a fluorescent surface. Push the
device to the front until the measurement result appears in the LCD. The
measurement value of the currently selected color channel is displayed.