5.
Fix the stirrer mount to the water jacketed referencing stage and tighten the screw using the M4
wrench.
Figure 39 Fixing the stirrer mount
Troubleshooting
When a magnetic bar does not move
1. Check whether the control unit is powered on.
2. Check the connection status between the control unit and the stirrer control cable.
3. Check whether the rotating speed (rpm) is properly set in the control unit.
When a Magnetic Bar always rotates irregularly
It is unavoidable that magnetic bars are aged with time, for example through sterilization, at excessive
temperatures, or mechanical stress (such as when you drop it).
1. Replace the magnetic bar with a new one.