Questions and Answers

  • C
    Cathy ThomasAug 11, 2025
    What to do if my Pfeiffer Vacuum Measuring Instruments show a continuous measuring signal of approximately 0 V?
    • A
      Amanda WilliamsAug 11, 2025
      If your Pfeiffer Vacuum Measuring Instruments display a continuous measuring signal of approximately 0 V, it could be due to several reasons. First, ensure the sensor connection is secure and that the supply voltage is properly established. Next, inspect the measurement cable for any defects or incorrect connections. Finally, the gauge might be in an undefined state; try switching it off, waiting for 5 seconds, and then switching it back on to reset it.
  • R
    Robert BrennanAug 17, 2025
    What to do if my Pfeiffer Vacuum PBR 260 display no signal?
    • W
      William HoldenAug 17, 2025
      If your Pfeiffer Vacuum Measuring Instruments are displaying no signal, there are a couple of things you can check. First, the internal data connection might not be OK, so try switching the gauge off, waiting 5 seconds, and then switching it back on to reset it. If that doesn't work, consider replacing the electronic unit.
  • D
    David MartinAug 22, 2025
    What to do if my Pfeiffer Vacuum PBR 260 Measuring Instruments show a measuring signal of approximately 0.5 V?
    • M
      martinbradleyAug 23, 2025
      If your Pfeiffer Vacuum Measuring Instruments show a measuring signal of approximately 0.5 V, it could be due to a Pirani error, indicating a defective measurement system, in which case replace the measurement system. Alternatively, the electronic unit might not be correctly connected to the measurement system; check the connection to resolve this.
  • K
    Kelli MillerAug 29, 2025
    Why isn't my Pfeiffer Vacuum PBR 260 Measuring Instruments switching over to BA measurement at low pressures?
    • L
      Lee AnthonyAug 29, 2025
      If your Pfeiffer Vacuum Measuring Instruments are not switching over to BA measurement at low pressures, it's likely that the Pirani zero point is out of tolerance. You should carry out a zero point calibration to resolve this issue.
  • J
    John ElliottSep 4, 2025
    What to do if my Pfeiffer Vacuum PBR 260 show a measuring signal of approximately 0.3 V?
    • K
      katherine41Sep 4, 2025
      If your Pfeiffer Vacuum Measuring Instruments show a measuring signal of approximately 0.3 V, it indicates a hot cathode malfunction, meaning the measurement system is defective. Replace the measurement system to fix this.