Measure spatial uniformity by scanning the uniformity section. Observe the trends
above and below the central mean value of a CT number prole for one or several rows or
columns of pixels as shown above.
Select a prole which runs from one side of the uniformity module to the opposite side.
Due to scanner boundary effects, typical proles start 2cm from the edge of the test
module.
Integral uniformity may be measured by determining the minimum and maximum CT
values along the prole and by using the following equation :
Integral Non-Uniformity = CTmax - CTmin
CTmax + CTmin
The phenomenon of “cupping” or “capping” of the CT number may indicate the need for
recalibration.