9
FREEZER COMPONENTS
1. Outer door: To open the door, grip the handle. When closing, lock the door latch completely.
2. Door latch: Always lock the latch when the outer door is closed.
3. Magnetic door gasket: This provides a tight door seal and prevents cold air leak. Keep clean.
4. Access port (back and bottom): This is used for leading a cable and sensor of a measuring
equipment, or nozzle of backup cooling kit to chamber.
5. Inner door: The operation of the inner door should be quick to minimize the temperature rise in
chamber. Lock the door latch completely when the door is closed. The door is removable for cleaning
or defrosting. See page 23 “Routine maintenance”.
6. Lock: Turn clockwise to 180
o
with a key and the outer door is securely locked.
7. Control panel (on the upper front of the door): Used for temperature setting and indication of
operating status is displayed on the panel. See page 10 for details.
8. Caster: 4 casters are provided to facilitate moving of the cabinet. For the installation, adjust the
leveling foot so that the front 2 casters cannot contact with the floor.
9. Leveling foot: These are screw bolts used to install and fix the unit. Adjust the height of the leveling
feet by turning the screw bolts until 2 front casters are away from the floor.
10. Fixture (on back side): 2 fixtures are provided as spacers between the cabinet and wall and also
serve as hooks to fix the unit. See page 12 “Installation”.
11. Air intake vent (grille): Do not block this vent to keep the proper cooling performance.
12. Power switch: This is for turning ON/OFF the power to the unit. ON – “I” OFF – “
○”
13. Space for temperature recorder: A temperature recorder (optional component) can be attached
here. See page 31 “Temperature recorder”.
14. Condenser filter (behind the grille): This filter prevents the dust from accumulating on the
condenser. The dusty filter may cause failure of refrigerating device. Clean the filter once a month.
See page 22 “Routine maintenance” for the cleaning.
15. Battery switch: This is a switch for a battery for power failure alarm. Normally, turn on this switch.
Be sure to turn off this switch if the freezer is not in operating for the long period (over one month).
16. Remote alarm terminal: This is used to notice an alarm condition of the unit to remote location.
Refer to page 17 “Remote alarm terminal”.
17. Space for optional component: