TROUBLESHOOTING
If the unit malfunctions, check out the following before calling for service.
Malfunction Check/Remedy
The chamber is not cooled
at all
The circuit breaker of power source is active.
The voltage is too low. (The status code “--2” is displayed. See
page 18.) In this case, call an electrician.
The power switch is not ON.
The cooling is poor The large amount of articles (load) is stored in the chamber at
one time. (The status code “--3” is displayed. See page 18.)
The ambient temperature is too high. (The status code “--1” is
displayed. See page 18.)
The latch of inner door is not closed completely. The outer door
is not closed firmly.
(The frost or ice between the cabinet and door gasket possibly
prevents door seal.)
The air intake vent is blocked.
The condenser filter is clogged. Always clean the filter when the
filter check lamp is lit.
The set temperature is not inputted properly.
The freezer is not away from the direct sunlight.
There is any heating source near the freezer.
A rubber cap and insulation for the access port are not set
correctly.
You put too many unfrozen articles into the freezer compartment.
Alarm test key cannot
activate the alarm
The alarm is activated only when the power switch is ON.
When only the buzzer or only the alarm is actuated by the alarm
test key, the unactuated part is out of order, and must be replaced.
Noise
The freezer is not installed on the sturdy floor.
There is anything touching the frame.
The freezer is not leveled with the leveling feet.
The freezer is in the status immediately after start up.
The unit sometimes causes a noise when the chamber temperature is
high due to the large load. The noise gets less and less accompanying
with the cooling of the chamber.
Back-up test switch does
not operate normally (when
the backup cooling kit is
installed)
The liquid carbon dioxide tank is empty.
The valve of the carbon dioxide tank is not opened.
The ambient temperature is too high. In this case, move the tank
to a cool location.
Inquire at liquid carbon dioxide suppliers about its check,
adjustment, installation, or move.