9
FREEZER COMPONENTS
1. Door: To open the door, grip the handle. When closing, lock the door latch completely.
2. Handle: Always grip the handle to open the door.
3. Lock: Turn clockwise to 180
o
with a key and the door is securely locked.
4. Defrost water vessel & storage container: The container can be used to collect the defrosted water
when defrosting.
5. Space for temperature recorder: A temperature recorder (optional component) can be attached here.
See page 23 “Temperature recorder”.
6. Leveling foot: These are screw bolts used to install and fix the unit. Adjust the height of the leveling
feet by turning the screw bolts until 2 front casters are away from the floor.
See page 12 “INSTALLATION”.
7. Caster: 4 casters are provided to facilitate moving of the cabinet. For the installation, adjust the
leveling feet so that the front 2 casters cannot contact with the floor.
8. Door latch: To lock the door, turn this latch downward. A padlock is also available.
9. Storage container: Made of styrol resin. Be careful not to damage the storage container by a
scraper at the time of defrosting.
10. Control panel: To display the temperature setting and running condition. See page 10 for details.
11. Remote alarm terminal: Used to notify an alarm condition of the unit to remote location. See page
17 for details.
12. Battery switch: Switch for battery used for power failure alarm. Always keep “ON”. Turn the
switch “OFF” when the unit is in no use for a long period (more than 1 month).
13. Fixture (on back side): 2 fixtures are provided as spacers between the cabinet and wall and also
serve as hooks to fix the unit. See page 12 “Installation”.
14. Power switch: This is for turning ON/OFF the power to the unit. This switch also activates as an
over-current breaker (15 A).
15. Access port: This is used for leading a cable and sensor of a measuring equipment.
Note: The door does not open quickly because the inside of the chamber becomes negative pressure
after the door is closed.