6CP30/15-3
2-3
CONTROLS AND INDICATORS
Control Panel (Cont’d)
14. Measurement Divider. Setting of this switch determines the voltage measurement divider ratio. This
switch must be set to match the actual high voltage output connection.
WARNING! If a test specimen intended to be tested at 10kV is connected to the 30kV output and the
Measurement Divider switch is set to 15kV, the actual voltage applied to the specimen will be
approximately 2 times the indicated voltage on the panel meter and an over-voltage may be applied to the
test specimen.
15. CM Range. Turn to select the desired current measurement range.
16. Start/Reset. Press to start the timer after test voltage has been reached. Press again to reset the timer to
the original setting.
17. HV Off Bypass. Press to allow high voltage to remain on after timer has expired. Indicator lights when
depressed.
18. Timer. Select desired test time using buttons on face of timer. See manual attachment.