6CP30/15-3
3-2
ELECTRICAL SET-UP
CONNECTIONS AND GROUNDING (QUICK REFERENCE)
Position set in the desired location.
Ground set by use of ground terminal on base of high voltage transformer.
Connect TX1 and TX2 cables between controls and high voltage transformer.
Connect SX1 - external interlock circuit as appropriate.
Connect return lead, provided, to low potential side of test specimen. See Ground-Guard-Return
connection instructions on following pages.
Connect high potential side of test specimen to either the 15kV (side) or 30kV (top) high voltage output.
Connect main power cord to appropriate receptacle.
(Further instructions on test connections may be found in ANSI C37.60)