17 Monitoring Noninvasive Blood Pressure
146
The NBP timestamp will normally show the completion time of the NBP measurement. Only under
the following conditions the timestamp shows the beginning of the measurement:
•when in
Auto or Sequence mode, and
the monitor is configured to synchronize the measurements in a measurement series to an "easy-
to-document" time. For example, if you start the first measurement at 08:23, and the
Repeat Time
is set to 10 minutes, the monitor automatically performs the next measurement at 8:30, then 8:40
and so on.
During Measurements
The cuff pressure is displayed instead of the units and the repeat time. An early systolic value gives you
a preliminary indication of the systolic blood pressure during measurement.
Starting and Stopping Measurements
Use the Setup menu or the SmartKeys to start and stop measurements.
Enabling Automatic Mode and Setting Repetition
Time
1 In the Setup NBP menu, select Mode.
2 Toggle between Auto and Manual, if necessary, to pick the measurement method.
3 If making an automatic measurement, select Repeat Time or press the Repeat Time SmartKey
and set the time interval between two measurements.
Action to be performed Setup menu SmartKeys
Start/Stop manual measurement
Start Auto series
Stop current automatic measurement
Start/Stop
Start/ Stop
Start manual measurement
Start Auto series
-
Start
Stop manual measurement
Stop current automatic measurement
-
Stop
Stop automatic, or manual measurement AND series Stop All NBP
Stop All