Report Recording. Choose from:
Manual - press the Record Report pop up key to trigger a manual request.
After Recorder Stop - report is recorded as soon as recorder becomes idle
Immediately - If a realtime recording is running, the monitor pauses it. The recording is
continued after the report has been recorded.
Average short term variability (STV) value is documented in [bpm] and [ms] if STV is configured as
part of the NST Report. This parameter is not considered as reassuring criteria.
NST Report Status Window
The NST Report window displays a detailed overview of the current NST status for any available
ultrasound fetal heart rate measurement. You can see:
NST Status - whether it is ready, ongoing, or the time and date at which it was stopped, or at which
it finished.
Runtime - the time that has elapsed since the NST began.
Accelerations - the number of FHR accelerations detected so far.
Baseline - the average baseline value.
Variability - the average variability value.
Short Term Variability - the current short term variability (STV) value.
Decelerations - the number of FHR decelerations detected so far.
FHR Availability - current statistical FHR availability value.
Sinusoidal - the current status of sinusoidal rhythm detection.
For criteria not yet met, a white arrow symbol marks the overall status on the top line, and also appear
against every criterion not yet met. A yellow symbol indicates detection of severe or prolonged
decelerations.
The pop-up keys let you perform the following actions:
FHR1, FHR2, FHR3 - switch to the window showing the current NST status for the fetal heart rate.
Record Report - prints the NST Report on paper.
Record Trace - records the trace episode that belongs to the current report. Depending on device
usage, the trace recording might be incomplete.
Setup - opens the Setup NST Report window.