8Non-Stress Test Timer
The non-stress test (NST) timer shows the elapsed time for the non-stress test. The timer counts up to
the time you set for the NST.
Setting NST Autostart/Autostop
You can set the recorder so that it starts automatically (NST Autostart) when the NST timer is started,
and stops automatically (
NST Autostop) when the NST is complete (when the set run time has
elapsed). As default,
NST Autostart is On, and NST Autostop is Off.
Viewing the NST Timer
You can configure the timer notification symbol ( ), the NST label, a progress bar and the
elapsed time to be displayed in the top left-hand corner of the screen. By default, the NST timer is not
displayed on the screen.
Alternatively, you can view the timer in the
Timers window.
To open the
Timers window:
Either
a. Press the
Timer SmartKey .
Or
b. Access the NST pop-up keys (see “Accessing the NST Setup Pop-up Keys” on page 88), and press
the
Timers key.
Timer Expiry Notification
When the timer expires, the color changes from blue to green, you hear a single tone, and a message
appears in the status line on the Main Screen.
The volume of the tone can be set in Configuration Mode.