5 Alarms
67
Select the Alarms Off SmartKey.
Or
1 Select Main Setup.
2 Select Alarms.
3 Select Alarms Off.
To Switch Individual Measurement Alarms On or Off
This applies to alarm mode All.
1 Select the measurement numeric to enter its setup menu.
2 Select Alarms to toggle between On and Off.
The alarms off symbol is shown beside the measurement numeric.
While Alarms are Paused or Off
• In the alarm field, the monitor displays the message Alarms Paused or Alarms Off, together with
the alarms paused symbol and the remaining pause time in minutes and seconds, or
alarms off symbol .
• No alarms are sounded and no alarm messages are shown.
• INOP messages are shown but no INOP tones are sounded.
The following INOPs are the only exceptions:
NBP CUFF OVERPRESS, Batt EMPTY and Batt MALFUNCTION (these INOPs are issued even if alarms
are paused or off).
If a disconnect INOP is present and alarms are paused or switched off, the measurement in question is
switched off.
Restarting Paused Alarms
To manually switch on alarm indication again after a pause, select the SmartKey Pause Alarms (or
Alarms Off) again.
Alarm indication starts again automatically after the pause period expires. If the monitor is configured
to stay paused infinitely, you must select
Alarms Off again to restart alarm indication.