6 Patient Alarms and INOPs
75
** Pulse HIGH SpO
2
The pulse rate has exceeded the high
alarm limit.
Numeric flashes and high
limit is highlighted, yellow
alarm message, alarm tone.
** Pulse LOW SpO
2
The pulse rate has dropped below
the low alarm limit.
Numeric flashes and low limit
is highlighted, yellow alarm
message, alarm tone.
** SpO₂ HIGH SpO
2
The arterial oxygen saturation has
exceeded the high alarm limit.
Numeric flashes and high
limit is highlighted, yellow
alarm message, alarm tone.
** SpO₂ LOW SpO
2
The arterial oxygen saturation has
fallen below the low alarm limit.
Numeric flashes and low limit
is highlighted, yellow alarm
message, alarm tone.
*** TACHY (Pulse)
***TACHY/P xxx>yy
SpO
2
The heart rate from the Pulse signal
has exceeded the tachycardia limit.
xxx denotes the highest measured
value, and yyy is the tachycardia limit.
Numeric flashes, alarm limit is
highlighted, red alarm
message, alarm tone.
Alarm Message From Condition Indication