6 Patient Alarms and INOPs
81
Toco UNPLUGGED
INOP tone
Toco Reconnect the Toco transducer to the monitor. Check all
connections are sound.
Unsupported LAN
INOP tone
Monitor There is a problem with the communication to the network and
central monitoring is currently not possible. Check the
connection.If the INOP persists, switch off the monitor and
contact your service personnel.
User I/F Malfunct.
INOP tone
Monitor Perform a visual and functional check of all the monitor input
devices. Contact your service personnel.
WRONG PAPER SCALE
INOP tone
Recorder The grid scale of the paper in the monitor does not match the
grid scale configured in the monitor. Make sure that you use the
correct paper and scale for your institution:
pre-printed: 30-240 in US and Canada, 50-210 in other
geographies.
INOP Message, Indication Source What to do