EverFlo / EverFlo Q User Manual
Yellow LED illuminates
continuously. The Red LED is
blinking and the Audible Alarm is
beeping periodically.
The device has
detected an impeded
oxygen ow condition.
Follow the troubleshooting guide
on the next page. Connect to a
back up oxygen source and call
your home care provider if your
troubleshooting actions fail to end
this alert condition.
Note: The No Oxygen Flow alarm
may take up to 1.5 minutes to
activate from the time the alarm
condition is present.
Yellow LED is blinking. The
Red LED is not illuminated and
the Audible Alarm is beeping
periodically.
The device has
detected a high
oxygen ow condition.
Follow the troubleshooting guide
on the next page. Connect to a
backup oxygen source, and call
your home care provider if your
troubleshooting actions fail to end
this alert condition.
Note: The High Oxygen Flow alarm
may take up to 1 minute to activate
from the time the alarm
condition is present.
Yellow LED illuminates
continuously. The Red LED is o
and the Audible Alarm is silent.
The device has
detected a low oxygen
condition (OPI units
only).
Continue using the unit but call
your home care provider about this
condition.
Note: The Low Oxygen alarm may
take up to 15 minutes to activate
from the time the alarm condition
is present.