EasyManuals Logo

Philips IntelliVue User Manual

Philips IntelliVue
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
Transmitter Non-RF Application Problems
8
4. If the problem is not solved by Steps 1 - 3, then
there may be a problem with the leadset switches
not being detected properly.
a. Check where the leadset attaches for dirt and clean
as necessary. Leadset switches are located next to
the reference and chest (standard ECG) or
reference and ā€œEā€ (EASI) lead wires.
b. The 3/5 lead switch may not be connected to the
ECG PCB properly.
c. Replace the Front-End Assembly
d. Replace the Main PCB.
NO SIGNAL INOP and an RF OUT OF LOCK
INOP at Telemetry Service Tool or Wave
Viewer
This INOP means that the transmitter has determined
that the phase-lock loop in the transmitter is no longer
functioning (This condition also generates a
NO SIGNAL
INOP at the Central Station). Replace the Main PCB.
quickref.fm Page 8 Wednesday, June 5, 2002 3:08 PM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue and is the answer not in the manual?

Philips IntelliVue Specifications

General IconGeneral
Battery BackupYes
Power Supply100-240 VAC, 50/60 Hz
CategoryPatient Monitor
ManufacturerPhilips
ModelIntelliVue
Display Size12.1 inches
Display TypeTFT LCD
ECGYes
SpO2Yes
NIBPYes
TemperatureYes
RespirationYes
Invasive Blood PressureYes
Cardiac OutputYes
EtCO2Yes
ConnectivityEthernet, USB, Serial
NetworkingYes

Related product manuals