Patient Monitoring
Service Guide
M1013A / M1019A IntelliVue G1/G5
IntelliVue G1/G5
M1013A/M1019A

Table of Contents

Questions and Answers

  • M
    mendozachristineAug 19, 2025
    What to do if Philips M1013A IntelliVue G1 Measuring Instruments show -?-, INOP tone, and blinking Setup LED?
    • S
      Sharon LaraAug 19, 2025
      If your Philips Gas Analyzer displays '-?-' instead of numerics, emits an INOP tone, and the Setup LED is blinking, this indicates a hardware problem. Check the connection to the monitor. Try switching the Gas Analyzer off and then on again to resolve the issue.
  • R
    rodriguezemilyAug 21, 2025
    What to do if Philips Measuring Instruments show GM OCCLUSION error?
    • M
      Michael BrownAug 21, 2025
      If your Philips Measuring Instruments display 'GM OCCLUSION' and numerics are replaced by '-?-' with an INOP tone, ensure the sample and exhaust lines aren't kinked. Check the airway adapter for water buildup, empty it, and reposition the adapter so the port faces upwards. Consider replacing the sample line, watertrap, or exhaust line.
  • P
    Peter DodsonAug 23, 2025
    Why Philips M1013A IntelliVue G1 show GM NOT AVAILABLE error?
    • P
      powelladamAug 23, 2025
      If your Philips Measuring Instruments display 'GM NOT AVAILABLE' with an INOP tone, it means the Gas Analyzer is either disconnected or switched off. Ensure it is properly connected and powered on.
  • G
    greenekathyAug 26, 2025
    How to fix Philips Measuring Instruments GM ZERO FAILED error?
    • D
      danielcarterAug 26, 2025
      If your Philips Measuring Instruments display 'GM ZERO FAILED' and numerics are shown as '-?-', check the exhaust tube for any occlusions or kinking, and replace it if necessary. You can also manually start another zero calibration.
  • G
    gtorresAug 27, 2025
    Why is Philips Measuring Instruments GM accuracy reduced?
    • W
      William JohnsonAug 27, 2025
      If your Philips gas analyzer displays numerics as '-?-' and the measurement accuracy seems reduced, check that the gas inlet, watertrap, and gas outlet tubing are not occluded.
  • J
    Jonathan DiazAug 29, 2025
    What does GM COMPONENT MALF mean on Philips M1013A IntelliVue G1 Measuring Instruments?
    • A
      adam93Aug 29, 2025
      If your Philips Measuring Instruments display 'GM COMPONENT MALF' and numerics are shown as '-?-', it indicates a gas analyzer component is malfunctioning. Try switching the gas analyzer off and then on again.
  • N
    Nancy ChaseSep 1, 2025
    What to do if Philips Measuring Instruments show the same value for et and in numerics?
    • A
      Alan GarciaSep 1, 2025
      If your Philips Measuring Instruments display the same value for both 'et' and 'in' numerics with a 'GM NO BREATH' message, check the patient connections to ensure they are correctly set up.
  • M
    Mark PotterSep 3, 2025
    Why Philips M1013A IntelliVue G1 show GM WARMUP error?
    • C
      Carla AdamsSep 3, 2025
      If your Philips Measuring Instruments display 'GM WARMUP' and numerics are shown as '-?-', it means the gas analyzer has not yet reached its operating temperature. Therefore, the measurement accuracy may be reduced.
  • A
    Alicia JonesSep 5, 2025
    What does AGT MEAS MALFUNCTION mean on Philips Measuring Instruments?
    • J
      Justin HallSep 5, 2025
      If your Philips Measuring Instruments display 'AGT MEAS MALFUNCTION,' with numerics replaced by '-?-' and an INOP tone, switch the gas analyzer off and then on again.
  • W
    warnerrebeccaSep 7, 2025
    What does GM CHECK WATERTRAP mean on Philips M1013A IntelliVue G1?
    • S
      Stephanie HenrySep 7, 2025
      If your Philips Measuring Instruments display 'GM CHECK WATERTRAP', check that the sample line and/or watertrap is not disconnected.

Summary

2 Installation and Patient Safety

3 Software Uploads

Checking Unit Functionality

Steps to verify IntelliVue G1/G5 functionality using the VISIA tool before software update.

Uploading Software Procedure

Step-by-step guide for updating the IntelliVue G1/G5 software using the VISIA tool.

4 Testing and Maintenance

Safety Tests Overview

Details safety test types, precautions, and general procedures for testing.

Electrical Safety Tests

Specific tests for protective earth resistance and equipment leakage current.

Performance Assurance Tests

Procedures for checking and calibrating the gas analyzer's performance.

5 Troubleshooting the Gas Analyzer

Technical Alarm Messages (INOPs)

Lists INOP messages, their indications, and recommended actions to resolve them.

6 Repairing the Gas Analyzer